1.0 OBJECTIVE:
To
lay down the calibration procedure of Digital Colony Counter to provide a system to calibrate, control
and maintain the equipment used for inspection measuring and test equipment in order
to ensure the correctness of measurement and test status.
2.0 SCOPE:
This
shall be applicable to the Digital
Colony Counter installed in Microbiology area of Quality Control Laboratory.
3.0 RESPONSIBILITY:
3.1
Technical
Assistant/ Helper.
3.2
Q.C Analyst/ Microbiologist.
4.0 ACCOUNTABILITY:
4.1
Manager-Quality
Control.
5.0 MATERIAL AND
EQUIPMENT:
5.1
Digital Colony Counter.
6.0 PROCEDURE:
6.1
Ensure that all
the connections of instrument are proper.
6.2
Operate the
instrument as per the operating instructions i.e. SOP .
6.3
Place an empty
petri plate on the illuminating surface; adjust the magnifying glass to view
the clear petri plate.
6.4
Press the tip of
marker pen 5 times till audio button indicates BEEP song and then check the
number displayed on screen, it should be ‘0005’.
6.5
Press the ‘RESET’
button, the digital display will show ‘0000’ again.
6.6
Press the marker
pen 10 times on surface of petri plate.
6.7
Repeat the same
procedure by pressing the marker pen 15 times, 20 times, 25 times, 50 times and
100 times.
6.8
Calibration of
colony counter is valid if the number displayed on screen is equal to the
number of time of marker pen pressed.
6.9
After completion
of work, press the ‘RESET’ button.
6.10 Switch
‘OFF’ the mains.
6.11 Precautions:
6.11.1 Do
not open the lid of petri dish.
6.11.2 Press
the marker pen on petri plates till the audio buzzer indicates beep song.
6.12 Acceptance
criteria:
6.12.1 Calibration
of digital colony counter is valid if the number displayed on screen is equal
to the number of time of marker pen pressed.
6.13 Frequency:
6.13.1 Once
in a month.
6.14 Record
the result on the “Calibration Record”,
“Calibration Log Book” & paste the “Calibration
Status Sticker” on Digital
Colony Counter.
7.0 REVISION LOG:
Revision No.
|
Effective Date
|
Reason
|
00
|
|
New
SOP
|
8.0 REFERENCES:
8.1 Not Applicable.
9.0 ANNEXURES:
Annexure 1:
Calibration record of Digital
Colony Counter.
Annexure:
1
Calibration
record of Digital Colony Counter.
CALIBRATION
RECORD OF DIGITAL COLONY
COUNTER
SOP Reference: _____________________
OBSERVATIONS
Remarks: The
colony counter is working satisfactorily/
not satisfactorily, requires/ does not require maintenance.
|
10.0
ABBREVIATIONS:
Abbreviation
|
Expanded Form
|
SOP
|
Standard operating
procedure
|
&
|
And
|
No.
|
Number
|
Ltd.
|
Limited
|
Q.C
|
Quality control
|
Q.A
|
Quality assurance
|
QCEB
|
Quality control
equipment calibration
|
F
|
Format
|