1.0 OBJECTIVE:
To
lay down the calibration procedure of Microscope.
2.0 SCOPE:
This
shall be applicable to the Microscope
of Microbiology Laboratory.
3.0 RESPONSIBILITY:
3.1
Technical
Assistant/ Helper.
3.2
Q.C Analyst/ Production
Pharmacist.
4.0 ACCOUNTABILITY:
4.1
Manager-Quality
Control/ Manager-Production.
5.0 MATERIAL AND
EQUIPMENT:
5.1
Microscope.
6.0 PROCEDURE:
6.1 Operate
the microscope according to the SOP No. BM/QCEO/SOP016-00.
6.2 Place
the stage micrometer (least count 0.01mm) on the slide holder stage.
6.3 Replace
the eye piece of microscope with the oculometer.
6.4 Fix
the first mark of oculometer with the first mark of stage micrometer using
different magnifying lens (10x, 40x and 100x).
6.5 Calculate
the least count oculometer using formula given below for each (x) of objective.
Least
count of oculometer (μm) = ______No. of marks of stage micrometer x 0.01
x1000_______
No.
of marks of oculometer matched with stage micrometer.
6.6 Record
result as per annexure-1.
6.7 Record
the result on the “Calibration Record”,
“Calibration Log Book” & paste the “Calibration
Status Sticker” on Microscope.
6.8 Frequency:
6.8.1
Yearly.
7.0 REVISION LOG:
Revision No.
|
Effective Date
|
Reason
|
00
|
|
New
SOP
|
8.0 REFERENCES:
8.1 Not Applicable.
9.0 ANNEXURES:
Annexure 1: Calibration
record of Microscope.
Annexure: 1
Calibration
record of Microscope
CALIBRATION
RECORD OF MICROSCOPE
Conclusion:
Satisfactory/ Not satisfactory.
|
10.0
ABBREVIATIONS:
Abbreviation
|
Expanded Form
|
SOP
|
Standard operating
procedure
|
&
|
And
|
No.
|
Number
|
Ltd.
|
Limited
|
Q.C
|
Quality control
|
Q.A
|
Quality assurance
|
QCEB
|
Quality control
equipment calibration
|
F
|
Format
|
mm
|
Millimeter
|
μm
|