Every compound, that is present in the nature, has a property to absorb, transmit, or reflect light (electromagnetic radiation) at a certain wavelength.
![]() |
Spectrophotometry is a technique which deals with the measurement of the interaction of light with materials |
Spectrophotometry is regarded as an instrumental
technique, based on the measurement of the absorption of electromagnetic
radiation in the ultraviolet (UV, 200–380 nm), visible (VIS,
380–780 nm), and near infrared region
Several
types of spectroscopic and spectrophotometric methods are applied to analyze
the samples. Among them, there are two primary methods which are highly
utilized;
Absorption
spectrophotometry,
which is based on the absorption of radiation at specific wavelength of light
to get absorption spectrum
UV-visible
spectrophotometry,
which is apprehensive with the reflectance of specific spectra of a given
material within UV and visible range of electromagnetic radiation spectrum
Principle
Standard operating procedure of taking absorbance or
transmittance reading on UV- Visible Spectrophotometer:
1.
OBJECTIVE
To
describe the procedure for analysis on UV-VIS Spectrophotometer
2.
SCOPE
This
procedure shall be applicable for analysis on UV-VIS Spectrophotometer
3.
RESPONSIBILITY
Officer/
Executive - Quality Control Dy. Manager - Quality Control
4.
ACCOUNTABILITY
Astt.
General Manager - QA/QC
PROCEDURE
- Ensure that the area is clean.
- Switch on the main power.
- Switch on the instrument.
- Allow stabilizing for 15 min.
- Set the desired wavelength with the knob in increasing order.
- Keep the sensitivity switch at the highest position.
- Put the mode selector at % T position.
- Set the filter wheel as per the wavelength.
- Adjust zero percent transmittance with the help of set zero control.
- Rinse reference cuvette and sample cuvette with blank solution 2 to 3 times.
- Adjust 100% T with the help of % knob.
- Remove the cuvette and drain out the blank solution from it. Rinse it with sample 2-3 times and filled with sample.
- Clean the outer surface of the reference cuvette with tissue paper.
- Place the cuvette into the sample holder and cover the cuvette compartment.
- The reading shown in the data position is the absorbance or % T of the sample.
- Record the results and make the entry in instrument usage log book.
- Remove both the cuvette and wash with purified water or methanol as required.
- Calibrate the instrument by running similar way and make an entry in the calibration record. Calibration Frequency: Monthly and after maintenance
5.
a
a
a